Sample rate
2 – 3.5ml/min
Tubing materials
Silicone and Tygon®
Required desk space for Instrument
50cm (H) x 45cm (W) x 45cm (D)*
For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size
Sugar Flame Photometer
51cm (H) x 39cm (W) x 35cm (D)
(20in x 15in x 14in)
Shipping
65cm (H) x 49cm (W) x 43cm (D)
(25in x 19in x 17in)
Weight
Sugar Flame Photometer – 14kg (30.8lbs)
Shipping – 25kg (55.1lbs)
Power requirements
100V – 250V AC at 50 or 60Hz
automatically selecting
Readout
LCD, four line, alpha numeric, back lit.
Optimal range
Single point calibration
Na 0.1 – 60ppm
K 0.05 – 100ppm
Multi Point Calibration
Na 0.1 – 1000ppm
K 0.05 – 1000ppm
Reproducibility
<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
Specificity
Na/K/Li = <1% to each other when
equal in concentration at <100ppm
Limit of detection (LOD) and
limit of quantification (LOQ)
LOD LOQ
Na – 0.03ppm Na – 0.1ppm
K – 0.02ppm K – 0.05ppm
Time to stability
Less than 15 seconds after sample is introduced into the flame
Ions measured
Potassium
Sodium
Interfaces
USB
4-20mA output capable of direct link to SCADA systems
Optional integrated printer
.CSV and .PDF generated reports and files via FP-PC software
Collection Cup providing 24/7 live automated sampling*
Recommended minimum warm up
time based on ambient temperature
21ºC – 25 minutes
Fuel requirement
Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of O.4l/min
*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.